Parker Hannifin
Hydraulic Filter Division Europe
FDHB500UK.
Specification
AutomaticParticleCounters(APC’s),havebeenwidely
used for many years in condition monitoring of hydraulic
fluids.However,itisonlyrecentlythatAPC’shavebecome
flexibleenoughtoenabletheinstrumentstobetakenoutof
thelaboratoryandusedon-lineinordertoobtainthemost
credibleformofresults.
Unusually,themovefromfixedlaboratoryuse,toportablefield
usehasnotbeenattheexpenseofaccuracyoruserflexibility,
buthasactuallyenabledtheinstrumentstobeusedovera
wider range of applications and situations.
ThemostcommonmonitoringtechniqueusedinAPC’sisthat
oflightobscurationorlightblockage.Here,afocusedlight
sourceisprojectedthroughamovingcolumnofoil,(inwhich
thecontaminantsbeingmeasuredarecontained),causingan
imageofthecontaminanttobeprojectedontoaphotodiode
cell,(changinglightintensitytoanelectricaloutput).
The electrical output of the photo diode cell will vary in
accordance with the size of the particles contained in the
columnofoil;thelargertheparticle,thebiggerthechangein
the photo diode electrical output.
On-lineAPC’smustbeabletotesttheoilsampleatwhatever
cleanlinessitisdeliveredtothemachine.Parkerthereforehad
todeveloptechnologytoensuretheon-lineAPCwasableto
testasamplewithouttheconventionallaboratorytechnique
whichrequiresdilution-apracticethatwouldhavebeen
simplyimpossiblewithaportableunit.
Bycarefuldesignandwindowsizing,gravimetriclevelsas
highas310mgofdirtperlitre,(equivalenttoupto4million
particles>6micronper100ml),canbeachievedwithout
makingtheinstrumentsusceptibletocountersaturation.
Thesehighsaturationpointon-lineAPC’s,whilstlosingnone
oftheaccuracyoftheirlaboratorycounterparts,enable
particlecountingtobecarriedoutquicklyandaccurately.
Core technology that proves itself in icountLaserCM
TheicountLaserCMportableparticlecounterfeatures
microprocessor controlled optical scanning for accurate
contaminantmeasurementwithacalibrationrangefromISO7
to ISO 22 with no counter saturation.
How does icountLaserCM work?
– Theparticlesaremeasuredbyaphotodiodethatconverts
light intensity to a voltage output which is recorded against
time.
As the particle moves across the window the amount of
light lost is proportional to the size of the particle. This
reduction in voltage is measured and recorded.
This “voltage” lost relates directly to the area of the particle
measured, is changed into a “positive” voltage and then in
turn changed into a capacitance value.
This value is counted and stored in the icountLaserCM
computerinoneof6channelsaccordingtoparticlesize.
– Readoutsaredisplayedonthehand-heldLCDinthe
accepted ISO and NAS standards ready for hard copy
printing or RS232 computer download.
– Theon-boardcomputerallowsstorageofupto300
test results.
A focused light source is
projected through a moving
column of oil.
Laser Optical Sensing
Parker Hannifin
HydraulicFilterDivisionEurope
FDHB500UK/icountLCM20